| Piezobalance Dust
Monitor Model 3521 / 3522 |
|
| Respirable Aerosol Mass Monitor
|
|
|
 |

|
| Impactors
for 10 and 4 micron cut off |
 |
Easy sensor
cleaning
|
|
|
| Applications |
- IAQ Investigations
- Industrial Hygiene
- Occupacianal Health & Safety
- Product Quality Control
- Laboratory Research
|
|
|
Features
|
- Data
logging ability allows user to log 500 samples.
- Simple cleaning mechanism
for easy maintenance.
- RS232C port available as
standard.
- Back-lit display.
|

|
|
Benefits
|
- While
conventional dust meters "count" particulates, the Piezobalance dust
meter "weighs" mass concentration of particulates.
- Data can be
reviewed on-screen, printed or downloaded to a computer.
- Software
included for easy data download and processing.
- Easy
operation requires no special training.
|
|
Principle
|
As
air sample enters the system, it travels through the impactor, which
captures and removes larger particulates away from the sample. Smaller
particulates become electrically charged and deposited on the
piezo-crystal. The total mass of the deposited particulates affects the
piezo-crystalfs frequency. Since the change in frequency is
proportional to the mass of the particulates, the actual weight of the
particulates is obtained.
|
 |
|
|
|
Specifications
|
Model
|
3521
|
3522
|
Measuring
Object
|
Airborn Particle Matter < 10 µm |
Airborn Particle Matter < 2.5 µm |
Impactor
|
10 µm, 5 µm |
2.5 µm |
Measuring
Range
|
0.01 to 10 mg/m3
|
Sampling
Flow Rate
|
1L/min
|
| Accuracy |
+/- 10% of reading +/- 1
digit (0 to 5 mg/m3)
+/- 20% of reading +/- 1
digit (5 to 10 mg/m3)
|
Data Logging
|
500
data and time stamped measurements |
Digital Output
|
RS-232C
|
Power Supply
|
AC or Ni-MH battery
|
| Dimensions |
H7.0h x
D5.9h x
W2.7h (180 x 150 x 65mm) |
Weight
(including batteries) |
4.4
lbs. (2 kg) |
|
|
KANOMAX
USA Copyright ©
2010 KANOMAX USA, INC. All Rights Reserved. |
 |
219 US
Hwy. 206, PO Box 372, Andover, NJ 07821
Tel: (973) 786-6386 / Fax: