PRINCIPLE:
Ultrafine airborne particles under 0.1um in cleanrooms cannot be detected by ordinary light-scattering particle counters.  To detect such small size particles, we have developed Condensation Nucleus Counter or CNC.  The CNC detects sub-0.1um particles by enlarging the particles with alcohol vapor.  The particles enlarge by going through supersaturated conditions formed by compulsion mixing of sampled air and high temperature alcohol vapor.

The CNC can measure and verify the particle levels within a short time, even in lean concentration environment in cleanrooms.  Thanks to its higher sampling rate.

 

 

                                     Structure of the Sensor

  Specifications

Type

Sensor-Counter Isolated Type

Principle

Vapor/Air Mixing Condensation Particle Counting

Maximum Condensation

103 Particles/cf

Sensitivity

0.05μm (100% Counting Efficiency)

Sampling Flow Rate

0.1 CFM (External vacuum source)

Alcohol

Propylene Glycol

Dimensions

280(W) × 400(H) × 150(D)mm

Power Supply

AC100V ±10%

   