Request Quote HDP5333

The HDP5333 Hydronic Differential Pressure Manometers are versatile, handheld, battery-operated manometers available in several basic ranges for positive or negative differential pressure measurement and can tolerate most liquid media compatible with 316LSS. The HDP5333 has a field adjustable damping, which allows the user to choose the level of display averaging rate corresponding to the fluctuation level common in many applications.

Hydronic meter kit includes: (2) six foot high-pressure rated tubing with shut off valves, high pressure 3-way valve, 8 ft poly tubing for purges, hard travel case, handheld carrying strap, 1/16″ x 1 1/2″, 1/8″ x 3″ and 1/8″ x 1-1/2″ pairs of pressure probes, 90° Flare to 1/4″ NPT fittings, adjustable 8″ wrench, PTFE Tape, and magnetic protective rubber boot. NIST Calibration Certificate comes standard.

  • Seven user-selectable English and metric units
  • Stores up to 40 readings for later recall
  • Backlight for use in dim areas
  • Digital dampening for low pressure – high resolution logging accuracy
  • Automatic resolution adjustment for finer control
  • HVAC system’s hydronic balancing
  • Designed to measure fluid pressure in hydronic systems
Service Compatible gases & liquids.
Wetted Materials Without valve: 316L SS, -3V option: 316LSS, Buna-N, silicone, grease, PTFE, brass 360, copper, reinforced acetal copolymer.
Accuracy ±0.5% FS, 60 to 78°F (15.6 to 25.6°C); ±1.5% FS from 32 to 60°F and 78 to 104°F (0 to 15.6°C and 25.6 to 40°C).
Pressure Hysteresis ±0.1% FS.
Pressure Limits See chart on catalog page.
Electronic Temperature Limits 32 to 140°F (0 to 60°C).
Storage Temperature Limits -4 to 176°F (-20 to 80°C).
Display 0.42″ (10.6 mm) 4 digit LCD.
Resolution See chart on catalog page.
Power Requirements 9 V alkaline battery, installed non-functional, user replaceable.
Process Connections Two 1/8″ (3.18 mm) female NPT.
Weight 14.1 oz (400 g).
Agency Approvals CE.